The AT100 is Based Upon the Magnetostrictive Principle. The Sensing Tube Contains a Wire, which is Pulsed at Fixed Time Intervals and the Interaction of the Current Pulse with the Magnetic Field Created by the Magnetic Float.
This Causes a Torsional Stress Wave to be Induced in the Wire.
This Torsion Propagates Along the Wire at a Known Velocity from the Position of the Magnetic Float and Toward Both Ends of the Wire.
A Patented Piezo-Magnetic Sensing Element Placed in the Transmitter Assembly Converts the Received Mechanical Torsion into an Electrical Return Pulse.
The Microprocessor-Based Electronics Measures the Elapsed Time Between the Start and Return Pulses and Converts it into a 4-20 mA Output, which is Proportional to the Level Being Measured.
SIL2 Certified IEC 61508*
High Accuracy: .01% of Full Scale
Superior Piezo Ceramic Sensor (Patent # 5,473,245)
Local Indication with LCD Display
Never Requires Re-Calibration: Set It & Forget It
Dual Compartment Housing with Separate Field Terminal Compartment
Loop Powered to 75’ (22M) Probe Length
Total and/or Interface Level Measurement
Pressure to 3000psig (207 bar), Std. 1800 psig (124.1 bar)
Temperature Range: -320 to 800º F (-196 to 427ºC) with options
Field Replaceable / Upgradable Electronics Module
Built In RFI / EMI Filter
Digital Communications
Housing type: Explosion Proof Powdered Coated Cast Aluminum or Stainless Steel, Dual Compartment
Electrical Connection: 1/2” FNPT or M20
Repeatability: 0.005% of full scale or 0.015”, whichever is greater
Non-Linearity: 0.01% of full scale or 0.035”, whichever is greater
Accuracy: 0.01% of full scale or 0.050”, whichever is greater
Supply Voltage: 13.5 to 36 VDC - Loop Powered; 9 to 32 VDC - Foundation Fieldbus
Reverse Polarity Protection: Diode in series with loop
Output/Communications: Standard 4-20 mA DC Loop HART protocol (standard) Foundation Fieldbus (optional) • ITK 5.1.0 Compliant • 5 AI and 1 PID blocks • 12.5 mA Quiescent Current Draw • LAS Capable Honeywell DE (optional)
Damping: Field adjustable by means of pushbuttons. Range: 0.1 to 36 seconds
Burnout: Jumper selectable upscale (21 mA) or downscale (3.6 mA)
Temperature: -40 to 170ºF (-40 to 77ºC) Ambient
Humidity: 0 to 100% R.H., non-condensing
Oil and Gas Industries
Petrochemical Industries
Chemical Industries
Power Generation Industries
Water and wastewater treatment
Food and Beverage Industries
Pharmaceutical Industries
Pulp and Paper Industries
Biotech Industries
Marine Industries