The role of humidity calibrations that are accurate, reproducible, and documentable is becoming more and more important.
The humidity calibrator HUMOR 20 developed by E+E is the ideal reference instrument for these requirements and is traceable to international standards.
The HUMOR 20 can be used in the humidity range of 10-95% RH both for monitoring cylindrical sensors and also for monitoring instruments with cubic dimensions.
Based on its operating principle, the HUMOR 20 can be used under typical conditions in a laboratory climate
.For operation HUMOR 20 requires only distilled water, filtered oil-free air with a pressure of 10 bar and a power supply between 90-230V AC.
The operation of the HUMOR 20 is based on a fundamental two-pressure process and thus is similar to instruments used in national bureaus for standards.
Accuracy : 0.3 - 0.9% RH
Supply : 90...230V AC
Operating resources: Distilled water, compressed air - filtered and free of oil with max. 10 bar
Operating range : 10-95% RH
Stabilisation time : < 3 min/measuring point
Calibration Laboratories
Reference Device
Bureau of standards
Manufacturers of measurement instruments