General information: With the Model CPH7600, Wally Box III, WIKA offers a further development of the well-known Wally Box I and Wally Box II.
The Wally Box III is a compact, portable pressure calibrator, which has been designed for mobile use as well as for stationary workshop and laboratory testing.
A special feature of this test instrument is that it is fitted with two independently-operating pressure systems. One pneumatic system is designed to be supplied from an external pressure supply, while the supply pressure of the second system is generated by the integrated electrical pressure generation.
This enables the user to perform different measurement tasks or calibrations, regardless of external pressure and power supplies.
Accuracy: The Wally Box III provides an accuracy of 0.025 % of reading ±0.01 % FS for the built-in, isolated pressure sensor.
The temperature compensation of the built-in sensor safeguards the accuracy for applications in the field. Readings can be displayed in one of 13 standard units.
Features: With the Wally Box III, not only accurate pressure measurements are possible, but also exact current and voltage measurements, as well as a DC 24 V supply for test items. Pressure values and electrical measured values are simultaneously shown in the display.
The calibrator is integrated into a robust, as well as dust and watertight transport case.
Thus the mobile pressure calibrator is particularly suited to use in harsh industrial environments.
A flow diagram on the front panel gives information on the connection of the individual pneumatic components. With the built-in NiMH rechargeable battery, a battery life of 50 hours or 125 pump cycles up to 20 bar is guaranteed.
The external universal power supply is designed for an operating voltage of AC 100 ... 240 V.
Certified accuracy: For each CPH7600 reference pressure sensor, the accuracy is certified by a factory calibration certificate which accompanies the instrument.
On request, we can provide a DKD/DAkkS calibration certificate for this instrument.
Measurement and electrical pressure generation at the touch of a button with integrated pump, from -0.8 ... +20 bar
Accuracy: 0.025 % of reading ±0,01 % FS (incl. calibration certificate)
Generation/measurement of 4 ... 20 mA, and 24 V voltage supply to power transmitters
Pressure supply via external compressed air line
Robust case design, IP 67
Calibration service companies and service industry
Measurement and control laboratories
Quality assurance